How to Calibrate a 67 MAX Infrared Thermometer

This article summarizes what you need to verify that your Fluke 67 MAX Infrared Thermometer is performing within its published specifications. Note that we are only talking about verifying the thermometer, not adjusting it. If your 67 MAX requires adjustment please contact your local Fluke service provider.

For more detailed instructions, please refer to the 67 MAX Infrared Thermometer Calibration Procedure, available on the Fluke website.

What equipment is required?

Verifying a 67 MAX thermometer is challenging due to the product specifications. Most infrared thermometer calibration instrumentation is designed for IR thermometers with much larger specifications and wider temperature range. The 67 MAX level of accuracy requires a special calibration setup. The following is a list of equipment we recommend for getting the job done.

4180 and 4181 IR Calibrators

  • Infrared black body source
    We recommend using a Fluke 4180 Precision Infrared Calibrator.

1504 Tweener Thermometer Readout

  • Reference thermometer readout
    You’ll use this device to accurately measure the temperature of the calibrator’s target, to improve the performance of your calibration beyond what the 4180 can do by itself. We recommend using a Fluke 1504 Tweener Thermometer Readout.

Probe

  • Reference thermometer
    We recommend using a Fluke 5610 Reference Temperature Probe. The probe works with the Tweener to make the reference measurement.

Required equipment and specifications

Classification

Minimum Use Specifications

Suggested Equipment

Black Body Source

Range:

Stability:

Resolution:

True Emissivity:

20 °C to 50 °C

0.1 °C

0.01 °C

> 0.9

Fluke 4180

Reference Thermometer Readout

Range:

Uncertainty:

5 kΩ to 10 kΩ 

100 µΩ/Ω (0.003 °C)

Fluke 1504 or better

Reference Thermometer

Sheath Diameter:

Sheath Length:

Range:

Uncertainty:

1/8 inch

6 or 9 inches

0 to 50 °C

0.01 °C calibration 0.01 °C long-term drift

Fluke 5610

Ambient thermometer

Uncertainty:

0.25 °C

Fluke 2626-S

 

Initial inspection and preparation

  • Inspect the 67 MAX thermometer.
    Check to ensure there is no external damage to the thermometer. It can be damaged by dropping, and damage might not be immediately apparent. While you’re doing this you might also wipe off any dirt and make sure the lens is clear.
  • Observe documented safety precautions for handling the calibration equipment.
  • Prepare the environment in which the equipment will be used. The location should:
    • Minimize reflected energy in the environment around the equipment. Eliminate anything greater than 30 °C that might reflect heat off the calibrator and into the thermometer field-of-view (for example, items like a coffee cup or even your hand can affect the measurement).
    • Not allow air to blow across the calibration equipment or the thermometer.

Equipment setup

Equipment setup for calibrating a 67 MAX thermometer

  • Set up the 2626-S to log ambient temperature in the calibration area.
  • Connect the 5610 probe to the 1504 readout.
    • Make sure the calibration coefficients that came with the probe are entered into the readout so it is displaying a calibrated temperature.
  • Place the probe inside the IR calibrator.
    • The 4180 has a hole in its top that allows you to put the probe in to measure the temperature of the target.
  • Configure the 4180.
    • Set the 4180 emissivity setting to 0.98 (which matches the emissivity setting of the 67 MAX).

      Position the 67 MAX thermometer to be calibrated

       

  • Position and configure the 67 MAX thermometer.
    • Ensure the thermometer has been in the calibration area for at least an hour to ensure its internal temperature is close to ambient temperature.
    • The thermometer should be 10 centimeters from the front plate of the 4180 and pointing so it is perpendicular to and focused on the center of the plate.
    • Turn on the 67 MAX by holding the trigger. Once a measurement is on the screen, press the SEL button so “body ON” is displayed. Press the SET button so “body OFF” is displayed. This turns off the skin-to-body temperature offset used in normal measurement mode and allows the 67 MAX to measure a blackbody correctly.

Taking the measurements

Test Points (°C)

Minimum No. of Samples (Reference Thermometer)

Minimum No. of Samples (DUT)

Maximum Reference Thermometer Std. Deviation Limit (1σ, °C)

Maximum DUT Std. Deviation Limit (1σ, °C)

Ambient Temperature Limit (°C) [Max – Min]

22

5

5

0.01

0.05

1

30

5

5

0.01

0.05

1

38

5

5

0.01

0.05

1

43

5

5

0.01

0.05

1

 

 

  • Take five samples of measurements each at a series of four test points, from the reference readout and from the 67 MAX.
    • We suggest you alternate, taking a reading from the reference, then taking a reading from the 67 MAX. Do that five times for each test point.
    • Record the maximum and minimum temperatures of the ambient condition observed during calibration. Ensure the difference between maximum and minimum is 1 °C or less.
  • Calculate the average and standard deviation from each set of the five measurements done at the four different temperatures.

Contact vs infrared temperature measurements

The 5610 and Tweener thermometer are contact measurement instruments. The 67 MAX thermometer is a non-contact infrared radiation measurement instrument. Contact and radiated temperature aren’t the same temperature. If you were using the 4180 in normal mode of operation you wouldn’t have to worry about this because it’s calibrated with radiation thermometry. However, normal mode of operation results in uncertainties that are too large to adequately verify the 67 MAX. When you bypass the normal calibration by using a contact thermometer, as described in this calibration instruction, you need to apply a correction to convert contact temperature to radiated temperature which is what the 67 Max sees..

To help with this, we have included a table of corrections that apply to this calibration. We have a technical note coming soon that will explain the math behind the correction but for the purpose of this article you can use the following table to identify the correction needed for each calibration point. Choose the correction that matches the observed ambient air temperature. (Note: It’s best to linearly interpolate between ambient temperature points to get the best correction value.)
 

 

Blackbody Contact Temperature

 

22 °C

30 °C

38 °C

43 °C

Ambient °C

Reference Thermometer Offsets (°C)

21

-0.05

-0.46

-0.84

-1.07

22

0.00

-0.41

-0.80

-1.03

23

0.05

-0.36

-0.75

-0.98

24

0.11

-0.31

-0.70

-0.94

25

0.16

-0.26

-0.65

-0.89

Note: The offsets listed in this table take into account the emissivity of the 4180 IR calibrator target surface and the ambient air temperature and assumes the 67 Max internal temperature is the same as ambient.

Here is an example of how to use the correction:

  • Ambient Temperature Reading = 23.0 °C
  • 4180 set to 38.0 °C
  • 1504/5610 Reference Temperature Reading = 38.000 °C
  • Corrected Reference Temperature = 38.000 °C + (-0.75) = 37.25 °C
  • Compare 37.25 °C with the 67 MAX displayed temperature

Calibration uncertainty

We have mentioned a few times that the purpose of this calibration setup is to get to a point where calibration uncertainties are adequate for verifying the 67 MAX. Calibration uncertainty is another complicated topic that will be covered in the upcoming technical note. Using good measurement techniques, described in other Fluke Calibration technical articles like those shown below, this setup can result in a calibration uncertainty of around 0.24 °C. We recommend applying the appropriate guardbanding practices to ensure the risk of false accept is adequate for your calibration program requirements.

Keep Learning

Accurate Fever Scanning with Infrared Forehead Thermometers: Issues, Solutions and How to Calibrate

Related Products

Fluke 4180 Precision Infrared Calibrator

Fluke 1504 Tweener Thermometer Readout.

Fluke 5610 Reference Temperature Probe

Mike Coleman, Director of Temperature Metrology at Fluke Calibration
Posted by Mike Coleman on Nov 17, 2020
Mike Coleman is Director of Temperature Metrology for Fluke Calibration. His career with Fluke and the former Hart Scientific spans 18 years and includes work in the Temperature Calibration Service Department and the Primary Temperature Calibration Laboratory, both in American Fork, Utah. Full Bio
  • Главная
  • Продукты
    • новые продукты
    • Калибраторы электрических
      • Электрические эталоны
      • Калибраторы электрических величин
      • Настольные мультиметры
      • Программное обеспечение для электрической калибровки
    • ВЧ-калибровка
      • Эталонные источники ВЧ-сигнала
      • Принадлежности для калибровки в ВЧ-диапазоне
      • Программное обеспечение для ВЧ-калибровки
    • Сбор данных и оборудование для испытаний
      • Настольные мультиметры
      • Сбор данных
      • Функциональные и параметрические генераторы сигналов
      • Программное обеспечение для сбора данных
    • Температурная калибровка
      • Первичные эталоны
        • Ячейки МТШ-90 с фиксированной точкой
        • Эталонные платиновые термометры сопротивления (ЭПТС)
        • Аппарат обслуживания
        • Сравнительный калибратор с жидким азотом
        • Мосты для измерения сопротивлений
        • Эталонные резисторы
      • Калибровочные ванны
        • Компактные калибровочные ванны
        • Эталонные поверочные термостаты
        • Ванны для специальных задач
        • Принадлежности калибровочных ванн
        • Контроллеры калибровочных ванн
        • Жидкости для калибровочных ванн
        • Калибровочные ванны по специальному заказу
      • Промышленные температурные калибраторы
        • Сухоблочные калибраторы для полевых условий
        • Сухоблочные калибраторы
        • Портативные сухоблочные калибраторы
        • Сухоблочные калибраторы для полевых условий
        • Микрованны
        • Инфракрасные калибраторы
        • Печи для термопар
        • Сухоблочный калибратор в виде двойного блока
        • Сухоблочный калибратор нулевой точки
      • Пробники/датчики
        • Платиновые термометры сопротивления (ПТС)
        • Термисторы
        • Термопары
      • Измерители температуры
      • Многофункциональные калибраторы
      • Программное обеспечение для калибровки температуры
    • калибровка влажности
      • Калибраторы влажности
      • Регистраторы и мониторы давления
    • Калибраторы давления
      • Поршневые манометры
        • Поршневые манометры NMI
        • Поршневые манометры абсолютного давления
        • Пневматические поршневые манометры высокого давления
        • Гидравлические поршневые манометры
        • Принадлежности для поршневых манометров
      • Контроллер/Калибраторы
        • Контроллеры/калибраторы низкого давления
        • Пневматические контроллеры/калибраторы давления
        • Пневматические контроллеры/калибраторы высокого давления
        • Гидравлические контроллеры/калибраторы
      • Мониторы давления
        • Мониторы низкого давления
        • Цифровые манометры
        • Мониторы эталонного давления
      • Грузопоршневые манометры
        • Пневматический грузопоршневой манометр
        • Масляные грузопоршневые манометры
        • Водные грузопоршневые манометры
        • Гидравлические грузопоршневые манометры высокого давления
        • Принадлежности для грузопоршневых манометров
      • Ручные калибраторы и мониторы
        • Калибраторы давления
        • Пневматическое управление давлением
        • Гидравлические компараторы/насосы давления
      • Handheld Pressure Calibrators
      • Калибровка данных для воздуха
      • Мониторы окружающей среды
      • Вспомогательные принадлежности для калибровки давления
      • Заказные системы калибровки давления
      • Программное обеспечение для калибровки давления
    • Калибраторы газовых расходомеров
      • Эталоны газового потока
      • Вспомогательные принадлежности для управления расходом газа
      • Первичный гравиметрический эталон расхода GFS
      • Программное обеспечение для калибровки потока
    • Калибраторы процессов
      • Калибраторы температуры
        • Ручные калибраторы температуры
        • Сухоблочные калибраторы и микрованны
        • Прецизионные цифровые термометры
        • Температурные зонды
        • Инфракрасные калибраторы
        • Гигротермометр с регистрацией данных
      • Калибраторы давления
        • Цифровые калибраторы давления
        • Ручные калибраторы давления
        • Грузопоршневые манометры
        • Прецизионные цифровые манометры
        • Ручные калибровочные насосы
      • Многофункциональные калибраторы
      • Калибраторы токовой петли
      • Программное обеспечение калибровки технологических процессов
    • Программное обеспечение для калибровки
      • ПО MET/CAL Software
      • Поддержка ПО MET/CAL
      • ПО сопровождения калибровки
      • Программное обеспечение для калибровки температуры
      • Программное обеспечение для калибровки давления
      • Программное обеспечение для калибровки потока
      • Программное обеспечение для механической/метрической калибровки
    • Сервис/Поддержка
    • Все приборы для калибровки
    • Переносные измерительные приборы
  • Покупателям
    • Где купить
    • Запрос цены
    • Запросить демонстрацию
    • Получить консультацию
    • Certified Pre-Owned Equipment
    • General Services Administration (GSA)
    • Financing Program
    • National Stock Numbers (NSNs)
    • Payment Options and Tax Info
  • Новости
    • Пресс-релизы
    • Акции и Продвижение
    • Metrology Salary Survey
  • Обучения/Мероприятия
    • Конференции и выставки
    • Курсы обучения
    • Встречи пользователей метрологического ПО
    • Вебинары
      • Интерактивные семинары
      • Архивные семинары
  • Литературы/Oбразования
    • Education Hub
    • About Calibration
    • Статьи и обучение
    • Литература, посвященная продукту
    • Руководства по продукту (руководства пользователя)
    • Видеоролики и виртуальные демонстрации
    • Blog
  • Поддержка/Сервис
    • On-line Service RMA
    • Программы обслуживания
    • База знаний
    • Аккредитация
    • Авторизованные центры обслуживания
    • Calibration Certificates
    • Community Forum
    • My MET/SUPPORT
      • Activate
      • Procedures
      • Software
      • Technical Bulletins
      • Priority Support
      • Manuals
    • Руководства по продукту (руководства пользователя)
    • Регистрация
    • Паспорта безопасности (SDS)
    • Safety, Service, and Product Notices
    • Загрузка программного обеспечения
  • О компании
    • Связаться с Нами
    • Посетите нас
    • Fluke Calibration
      • Научная калибровка температуры в жестких условиях
      • Калибровка давления и потока для инструментов DH
      • Калибровка давления для приборов измерения давления
      • Калибровка давления Ruska
    • Вакансии
    • Компании в составе Fluke
    • Корпоративные стандарты компании
    • Часто задаваемые вопросы
На главную|Blog|How to Calibrate a 67 MAX Infrared Thermometer
Fluke Calibration RU Feedback

Нам нужен ваш отзыв.
Вы можете помочь нам
улучшить этот веб-сайт.

Спасибо!

©1995-2021 Fluke Corporation
     

Secondary menu

  • Дисклеймер
  • Конфиденциальность
  • Веб-мастер
  • Terms of Use
  • Условия продажи